UPC5200/UPC5210 High Pressure Calibration Standard
Standard Features
- Available in enclosure with carrying handles (UPC5200)
- Two sets of ranges to choose from: 10,000/5,000/2,000 psi and 5,000/2,500/1000 psi
- Three independent selectable ranges in each instrument
- ±0.05% F.S. accuracy of each independent selectable range
- Available with Light Emitting Diode (LED) or Liquid Crystal Display (LCD)
- Pressure display resolution 0.02% F.S. typical for each range
- Different technicians can calibrate the same Device-Under-Test and get identical results
- Electronically controlled 10:1 intensifier system
- Easy to operate, training can be accomplished in 10 minutes
- Easy to carry; approximately 51 lb
- Adjustable overpressure protection above input pressure
- Precision pressure vernier control
- Voltage/current display mode for transducer testing
- 32 VDC excitation output
- Non-destructible, finger tip control micrometering valves for pressure control
- Automatic pushbutton calibration, microprocessor controlled and maintained
- Quick-disconnect fitting and hose included
- Pneumatic only; no hydraulics or distilled water messes
- Calibration certificate included; traceable to the National Institute of Standards and Technology(NIST)
Applications
- Any high-pressure, pneumatic-only application
- Initial test of high-pressure hydraulic systems
- Boiler feed pressure for power plants
Providing sophisticated performance in high-pressure applications, the durable UPC5200 high-pressure calibration standard is the most effective digital pressure calibration standard in its class.
The UPC5200 high-pressure calibration standard excels in applications presenting pressure levels up to 10,000 psi. Featuring automatic pushbutton calibration, this unit guarantees 0.05% accuracy in three switch-selectable ranges. The exclusive patented precision pressure Orion 3A vernier provides infinite control to obtain exact pressure readings, while a reliable overpressure regulator protects units under test from permanent damage.
The fully-pneumatic UPC5200 high-pressure digital calibration standard is powered by nitrogen and eliminates the mess and oil thermal equilibrium problems common to hand pump hydraulic systems. Clean operation makes this calibrator ideal for use in aerospace, food and beverage and general industry. The UPC5200 high-pressure calibration standard is easy to operate and training can be accomplished in as little as 10 minutes. Different technicians can calibrate the same transmitter and get identical results...every time.
For special high-pressure applications, the portable UPC5200 or 19-inch rack-mount UPC5210 provides the ultimate in dependable, high-precision performance.
Condec UPC5200/UPC5210 Model Numbering Guide
UPC5200/UPC5210 Specifications
Available Calibration Modes: Overall accuracy: Power: Temperature Range: Pressure Media: Regulated Pressure Gauge: Pressure Media Filter: |
Control Valves: Internal Piping: Output Test Port: Pressure Hoses: Vernier Control: Booster Pump Solenoid Valve: Accumulator Gauge: |
Input Pressure Regulator: UPC5200 Internal Pressure Sensor: External Carrying Case: Current Display: Weight: |
Parts
Part # | Description |
---|---|
61320 | Calibrator,UPC5200BAA Portable High Pressure Calibration Standard, Ranges 5000/2500/1000 PSI, Gage Only, Accuracy 0.05% FS of Selected Range, LED Display, Power Requirement 120 VAC, Includes National Institute of Standards & Technology (NIST) Calibration Cert, Test and Input Port Hoses, Power and Electrical Cables If written data points are required for new units additional charge. Note: Pressure Trap PN 58478 and Calibration Module PN 60109 Sold Separately. |
95586 | Calibrator,UPC5200CAA Portable High Pressure Calibration Standard, Ranges 10,000/5,000/2,000 PSI, Gage Only, Accuracy 0.05% FS of Selected Range, LED Display, Power Requirement 230 VAC, Includes NIST Cal Cert, Test and Input Port Hoses, Power and Electrical Cables. If written data points are required for new units additional charge. Note: Pressure Trap PN 58478 and Calibration Module PN 60109 Sold Separately. |
61239 | Calibrator,UPC5200AAB Portable High Pressure Calibration Standard, Ranges 10,000/5,000/2,000 PSI, Gage Only, Accuracy 0.05% FS of Selected Range, LCD Display, Power Requirement 120 VAC, National Institute of Standards & Technology (NIST) Calibration Certificate included, If written data points are required for new units additional charge. Note: Pressure Trap PN 58478 and Calibration Module PN 60109 Sold Separately. |
61322 | Calibrator,UPC5200BAB Portable High Pressure Calibration Standard, Ranges 5000/2500/1000 PSI, Gage Only, Accuracy 0.05% FS of Selected Range, LCD Display, Power Requirement 120 VAC, National Institute of Standards & Technology (NIST) Calibration Certificate included, If written data points are required for new units additional charge. Note: Pressure Trap PN 58478 and Calibration Module PN 60109 Sold Separately. |
109561 | Calibrator, UPC5200CAB Portable High Pressure Calibration Standard, Ranges 10,000/5,000/2,000 PSI, Gage Only, Accuracy 0.05% FS of Selected Range, LCD Display, Power Requirement 230 VAC, National Institute of Standards & Technology (NIST) Calibration Certificate included, If written data points are required for new units additional charge. Note: Pressure Trap PN 58478 and Calibration Module PN 60109 Sold Separately. |
104069 | Calibrator,UPC5200DAB Portable High Pressure Calibration Standard, Ranges 5000/2500/1000 PSI, Gage Only, Accuracy 0.05% FS of Selected Range, LCD Display, Power Requirement 230 VAC, Includes NIST Cal Cert Test and Input Port Hoses, Power and Electrical Cables. If written data points are required for new units additional charge. Note: Pressure Trap PN 58478 and Calibration Module PN 60109 Sold Separately. |
Options
Part # | Description |
---|---|
58478 | Pressure Trap,UPC5200 All Ranges, PIN7000 PIN7010 PIN8000 PIN8010 All, UPS3110 and DLR3110 Ranges 5000 PSI and Above Operation |
63259 | LIT,Manual UPC5200 Operating |
55542 | Fitting Kit,Test Port Standard quick-disconnect test port (output) fitting SST male plug on one end and 7/16-20 37 DEG AN-4 JIC male on opposite end w/ internal 40 micron filter, 11000 PSI working pressure, Requires, included, one Cheat Seal 54854 for UPC5200 UPC5210 PIN7000 PIN7010 PIN8000 PIN8010 UPS3110 DLR3110 |
151017 | Fitting Kit, Output Port Quick Disconnect Fitting Male Coupling Body, other end 1/4-18 NPT Female SST with 40 Micron Filter, 11000 PSI Working Pressure, for PIN8000 PIN8010 UPC5200 UPC5210 PIN7000 PIN7010 UPS3110 DLR3110 - NOTE: May be used to connect DPDG Gauge to Output Port. |
55280 | Hose Kit,Test Port Output UPC5200, PIN7000, PIN8000 5 Foot Long with Quick Disconnect SST Plug on One End and 1/4 inch 37 Degree Female AN Swivel Tube Coupling on Opposite End, 3/16 in Hose ID, 10,000 PSI (690 BAR) Working Pressure |
131096 | Hose Kit,Test Port Output 10 ft. test hose for UPC5200, UPS3110, DLR3110, PIN7000 & PIN8000 (5000 psi and above) Quick Disconnect SST Plug on One End and 1/4 inch 37 Degree Female AN Swivel Tube Coupling on Opposite End |
78939 | Hose ASSY,Pneumatic 5 ft Fill Port One End of Hose 7/16-20 JIC 37 Degree Female AN-4 Swivel Other End CGA-580 Type Nipple for Source 3000 65315 or 98917 or 131385 |
55366 | Hose Kit,Fill Port Input UPC5200 10 Foot Long with CGA-580 Male Nipple on One End and 1/4 inch 37 Degree Female AN Swivel Tube Coupling on Opposite End |
55369 | Hose Kit,Fill Port Input UPC5200 15 Foot Long with CGA-580 Male Nipple on One End and 1/4 inch 37 Degree Female AN Swivel Tube Coupling on Opposite End |
162561 | Hose Assy,Pneumatic 20 ft Fill Port One End of Hose 7/16-20 JIC 37 Degree Female AN-4 Swivel Other End CGA-580 Type Nipple, for Source 3000 PNs 65315, 98917, 131385 |
60109 | CONDEC Calibration Module For UPC5000 UPC5010 UPC5100 UPC5110 UPC5200 UPC5210 |
An ISO 9001 Registered Company ©2024 Rice Lake Weighing Systems